Friday, 20 January 2017

Number of location for Non viable particle count (NVPC) in Clean Room as per Area

 Deriving the number of sampling locations.
 Earlier the numbers of locations for monitoring of Non Viable Particle Counts was as per square root of the total area but now it has been revised and the numbers of the location has been fixed with respective of the area in square meter. If considered area falls in between the two values, then greater of the two values shall be selected

Area (m2) Less than or equal to
Minimum number of sample locations


Area (m2) Less than
or equal to
Minimum number of sample locations
2
1
76
15
4
2
104
16
6
3
108
17
8
4
116
18
10
5
148
19
24
6
156
20
28
7
192
21
32
8
232
22
36
9
276
23
52
10
352
24
56
11
436
25
64
12
636
26
68
13
1000
27
72
14
>1000
N = 27 X (area of clean room in m2 / 1000)             